2kV HV-SMU Board for 431-TT
12.25.2013 | Category, Products
431-TT:HV-SMU board for 2kV Application
1. GENERAL
This is the HV-SMU 2kV application board for model 431-TT.
The standard board is PT1324-000 that is able to measure the leakage current and breakdown voltage up to +2kV/+99.99mA specification.
-2kV/-99.99mA testing is optional and PT1325-000 board is required.
PT1325-000 board cannot be used without PT1324-000 as stand alone.
SMU method made possible to have voltage forcing, current forcing, voltage sensing & current sensing.
In order to have high voltage power forcing (±2.0kV), 2.4kV HV POWER UNIT is required.
2. SPECIFICATIONS
【2.4kV HV POWER UNIT】
Output voltage | : ±2.4kV |
Output numbers | : 2(System) |
【PT1324-000, PT1325-000】
Voltage range | : 0.000V~±2.000kV |
Current range | : 000.0nA~±99.99mA |
MPX | : 2 stations switching |
Function | : Waveform monitor |
The Auto Calibration function of CSU unit can be applied. |
For more detail information, please contact us.