431-TT:HV-SMU board for 2kV Application

1. GENERAL

This is the HV-SMU 2kV application board for model 431-TT.
The standard board is PT1324-000 that is able to measure the leakage current and breakdown voltage up to +2kV/+99.99mA specification.
-2kV/-99.99mA testing is optional and PT1325-000 board is required.
PT1325-000 board cannot be used without PT1324-000 as stand alone.
SMU method made possible to have voltage forcing, current forcing, voltage sensing & current sensing.
In order to have high voltage power forcing (±2.0kV), 2.4kV HV POWER UNIT is required.

 

2. SPECIFICATIONS

【2.4kV HV POWER UNIT】

Output voltage : ±2.4kV
Output numbers : 2(System)

 

【PT1324-000, PT1325-000】

Voltage range : 0.000V~±2.000kV
Current range : 000.0nA~±99.99mA
MPX : 2 stations switching
Function : Waveform monitor
The Auto Calibration function of CSU unit can be applied.

For more detail information, please contact us.